Lab-Tacular
Serial Dilution Prep Calculator
Enter the starting concentration, dilution factor, and number of steps to get the per-step concentration automatically. Add the total volume per tube to also get transfer and diluent volumes.

If provided, transfer and diluent volumes are calculated too.

The dilution factor must be greater than 1, and up to 12 steps can be calculated. Results update automatically as you type.

The dilution factor in the table can differ from step to step. Changing any of them switches the field above to "Custom".

How to use

  1. Enter the starting concentration, the dilution factor and the number of steps. Ten-fold across five steps means a factor of 10 and 5 steps.
  2. Total volume per tube is optional. Leave it blank and you get the concentration at each step; fill it in and the transfer volume and diluent volume are worked out as well.
  3. Editing a cell in the Dilution factor column lets a step differ from the rest. Change any one of them and the field above switches to ‘Custom’.
  4. [Same factor throughout] puts them back. An empty cell follows the shared factor above.
  5. Results update as you type. There is no calculate button.

Formula & notes

A serial dilution reaches a wide range of concentrations by repeating the same small dilution. Going ten-thousand-fold in one step means transferring a volume too small to pipette accurately; four ten-fold steps arrive at the same place with a workable volume every time.

Each step divides the previous concentration by the factor.

Concentration at step n
Cₙ = C₀ ÷ (factor)ⁿ

Where the factor differs from step to step, divide by the product of the factors up to that point.

With mixed factors
Cₙ = C₀ ÷ (f₁ × f₂ × … × fₙ)

Once the total volume V per tube is set, the transfer and the diluent follow from it.

Transfer
V ÷ factor
Diluent
V − (V ÷ factor)

A ten-fold dilution into 1 mL means carrying 100 μL from the previous tube and adding 900 μL of diluent.

Practical notes

  • Mix thoroughly at every step. Carrying an unmixed tube forward passes its error into every step after it, and because the errors of a serial dilution multiply, an early one costs the most.
  • Change the tip between steps. Concentrated solution clinging to the outside of a tip raises the concentrations downstream above the calculated values.
  • When the transfer falls below about 1 μL, a smaller factor over more steps is more accurate. Seven five-fold steps are often easier to handle than five ten-fold ones.
  • The volumes in the table make each tube up to the total volume you set. If you need more than that from the last tube, set a larger total — every tube loses what is carried to the next one.
  • Substances that adsorb readily — some proteins, oligonucleotides — can come out below the calculated value at the dilute end. Consider low-binding tubes or a carrier.

FAQs